channel. This will verify the mounting and interfacing characteristics of the
sensor, to ensure it falls within requirements. It should be inspected to ensure
that a proper pressure sample will be taken given the orientation of the sensor in
the flow channel. In addition, it should be verified that the sensor is not blocking
flow over or into the alcohol sensor.
After this is completed, a small amount of airflow from a human breath should be
directed into or onto the airflow or pressure detection mechanism. At this point,
any sort of result should be verified in the rest of the system. The approximate
same level of exhalation effort and amount should then be directed into the
mechanism again, and a similar result verified. Alternatively, a low pressure
mechanical pressure exertion device can be used, such as a low pressure can of
compressed air or perhaps a medical inhaler (which contains no chemicals). If
the results are indeed similar, this portion of the test is passed. Then, a deep
breath, similar to the one required for alcohol testing, should be performed on the
detection mechanism to ensure a proper reading is obtained, and for the proper
duration. This test should then be repeated two more times to ensure the sensor
produces consistent results within an acceptable range to prove its utility in the
application of deep lung exhalation detection.
Table 5.2.6-1 provides the results of repeated testing and evaluation of the
pressure detection system of the Voog handheld breathalyzer. Based on this
data, a threshold value was established to determine the strength and quantity of
breath flow necessary for a sufficient sample reading.
Differential Pressure Sensor Output
Input Sample
Reference Voltage
ADC Output
(1.46 mv/step)
No Breath
1.5V
750
Breath Sample 1
(minimal)
1.5V
723
Breath Sample 2
(less minimal)
1.5V
690
Breath Sample 3
(ideal)
1.5V
652
Breath Sample 4
(maximum)
1.5V
595
Table 5.2.2-1 Pressure Sensor Test Data
5.3 Software Verification Requirements
Much like the hardware components of a complete system design, the software
components must also have established requirements for the confirmation of